9.1 External Trigger Input
205
3
Chapter 9 Ext
ernal Cont
rol
9
The external control terminals on the instrument support trigger signal input and
output.
Be sure to read p. 52 to p. 53 for external control terminal connection details.
Triggering can be controlled by applying a signal from an external trigger source
(p. 136).
This allows synchronous operation of multiple instruments by parallel triggering
See "Using External Triggering"(p .136) for setting procedures and trigger signal
details.
The external signal input function can be changed to insert event marks.
See:
"Inserting Event Marks Using External Input Signals"(p .122)
External Control
Chapter 9
9.1 External Trigger Input
External signal
input
(EXT.TRIG)
Trigger event occurs.
2.5 ms or
greater
2.5 ms or greater
HIGH: 2.5 to 5.0 V
LOW: 0 to 0.8 V
Triggering occurs at the selecting rising or falling edge.
(according to the selected
[Ext trig]
detection edge)
Voltage range
HIGH level: 2.5 to 5.0 V, LOW level: 0 to 0.8 V
Pulse width
HIGH period: 2.5 ms or greater, LOW period 2.5 ms or greater
Maximum input voltage
DC0 to 10 V
Trigger Input Signals
Summary of Contents for LR8410-20
Page 1: ......
Page 2: ......
Page 12: ...Contents x...
Page 32: ...Operating Precautions 20...
Page 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Page 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Page 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Page 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Page 316: ...13 3 Disposing of the instrument 304...
Page 360: ...Index Index 4...
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......