5.1 Triggering Measurement Start and Stop
137
3
5
Chapter 5 S
p
ecif
ying Crite
ria
for M
easurement
s
When trigger timing is set to
[Start]
or to
[Start/Stop],
not only the
waveform following the trigger can be measured, but a specified
span of the waveform before triggering as well.
However, when trigger timing is set to
[Stop]
, pre-trigger settings are
disabled.
Make these settings on the
[Trig & Alm]
screen.
See:
"Key Setting Procedure" (p .130)
When You Want to Measure Data before Triggering (Pre-trigger)
T
[Pre-Trig]
Set the amount of time to record before each trigger
event.
To also measure the waveform after triggering, the recording
time must be set longer than the pre-trigger time.
(See "Pre-Trigger and Recording Times" below.)
The maximum settable
time is displayed.
Increment/decrement by 1
(when the recording interval is longer than 2 seconds,
increment/decrement by one recording interval)
Increment/decrement by 10
(when the recording interval is longer than 10 seconds,
increment/decrement by one recording interval)
The instrument pre-trigger time can be set up to 99 days. Logger Utility
settings of 100 days or more are reduced to 99 days when transferred to
the instrument.
Difference between
[
Waiting for pre-trigger
]
and
[
Waiting for trigger
]
When measurement is started, the specified pre-trigger length is recorded. This period is indicat-
ed as the
[Waiting for pre-trigger]
.
After the specified pre-trigger length has been recorded, the period indicated as
[Waiting for trig-
ger]
continues until a trigger occurs.
During the
[
Waiting for pre-trigger
]
interval, trigger criteria are ignored.
T
Pre-Trigger (15 min)
Recording
(10 min)
Measurement ends
with the trigger event
When the recording time is shorter
than pre-trigger time
When the recording time is longer than
pre-trigger time
T
Pre-Trigger (5 min)
Recording
(15 min)
Measuring before
and after triggering
Ends when the record-
ing time has elapsed
Pre-Trigger and Recording Times
Summary of Contents for LR8410-20
Page 1: ......
Page 2: ......
Page 12: ...Contents x...
Page 32: ...Operating Precautions 20...
Page 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Page 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Page 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Page 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Page 316: ...13 3 Disposing of the instrument 304...
Page 360: ...Index Index 4...
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......