6.2 Saving Data
166
During real-time saving, removable storage can be replaced without interrupting mea-
surement. This procedure describes replacing a USB flash drive.
Replacing Removable Storage During Real-Time Saving
Move the cursor to
[EJECT]
.
Execute
Insert a formatted removable
storage.
Measurement data acquired while the de-
vice was removed is automatically written.
See:
Formatting: (p. 62)
Insertion: (p. 60)
Remove the USB flash drive (p. 61).
Right bottom of the screen
Move the cursor to
[Unmount USB memory]
.
Execute
Move the cursor to
[Yes]
.
Execute
Execute
A "Replaceable." message appears.
When both a SD memory card and USB flash
drive are installed and either is removed, the save
destination automatically switches to the remain-
ing media.
• If a storage device is not replaced within ten minutes after selecting "Yes" in
the "Do you want to remove the USB memory/SD card?" confirmation dia-
log, data may overflow internal buffer memory and be lost.
• If a storage device is replaced during real-time saving, data is saved to a
new file.
• If measurement ends while the removable storage is ejected, even if a
removal storage is then inserted, the remaining data is lost. In that case,
when connected to a LAN, data can be read from the instrument via the
HTTP function. See "10.4 Remote Measurement with the HTTP Server
Function" (p. 236).
Summary of Contents for LR8410-20
Page 1: ......
Page 2: ......
Page 12: ...Contents x...
Page 32: ...Operating Precautions 20...
Page 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Page 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Page 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Page 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Page 316: ...13 3 Disposing of the instrument 304...
Page 360: ...Index Index 4...
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......