Appendix 12 Instrument/Unit Communications Interruptions and Reestablishment
A
39
Appendix
Start trigger rising slope
OR
Stop trigger falling slope
AND
Stop trigger rising slope
AND
Stop trigger
Start trigger
Connection
Connection interruption
Instrument data
Data following the stop trigger is not restored,
but rather reported as “NO DATA.”
Start trigger
Stop trigger
Example: Stop trigger criteria
When the stop trigger falling slope
AND and the stop trigger rising slope
AND are the same
Data before the start trigger is not saved.
①
②
③
④
Unit data
Start trigger rising slope
OR
Stop trigger falling slope
AND
Stop trigger rising slope
AND
When unit 1 and then unit 2 are connected to the instrument (in that order)
(1) Recording of data starts at the start trigger position.
(2) All data from unit 1 is restored.
(3) Data from unit 2 is restored.
(4) Since the stop trigger criteria were met while data from unit 2 was being restored, both
the data restoration and measurement stop.
Unit 1
Unit 2
Unit 1
Unit 2
Summary of Contents for LR8410-20
Page 1: ......
Page 2: ......
Page 12: ...Contents x...
Page 32: ...Operating Precautions 20...
Page 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Page 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Page 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Page 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Page 316: ...13 3 Disposing of the instrument 304...
Page 360: ...Index Index 4...
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......