50
_____________________________________________________________________
3.5 Trigger Functions
______________________________________________________________
Within the
period range
Within the
period range
Upper limit of
the period range
20 ms
20 ms
21 ms
22 ms
Reference level
0.000 V
Example for period Trigger
To measure a 50-Hz power trend waveform with period
deviation trigger conditions between 19 ms (about 52.6 Hz)
and 21 ms (about 47.6 Hz), set the following:
Lower limit: 19 ms Upper limit: 21 ms
Reference level: 0.000 V Slope: