99
_____________________________________________________________________
5.6 External Trigger Function
______________________________________________________________
DANGER
To avoid damaging the instrument, do not apply
voltage outside of the range -5 to +10 V to the external
trigger terminal.
External trigger
terminal
GND terminal
Procedure
Screen: CH SETTING
Flashing cursor
OFF
External trigger is disabled.
ON
External trigger is enabled.
5.6 External Trigger Function
・
An external signal can be used as trigger source.
・
The external trigger facility can be used to synchronize a
number of the 8715-01 units for parallel operation.
1. Press the CH SET key to display the CH SETTING
screen.
2. Move the flashing cursor to Ext Trig and use the
keys to make a setting.