3.5 Trigger Functions
______________________________________________________________
Flashing cursor
Flashing cursor
OFF
Trigger filter is disabled.
ON
Trigger filter is enabled. Filter width: 0.5 DIV
3. Move the flashing cursor to the position shown in the
figure below, and set the lower limit of the trigger level
using the
keys.
4. Move the flashing cursor to the upper limit, and set the
upper limit of the trigger level using the
keys.
The upper limits must not be smaller than the lower
limits, or the lower limits must not be larger than the
upper limits.
5. Move the flashing cursor to the position shown in the
figure below, and set the trigger filter using the
keys.
If the trigger condition has already occurred before
measurement starts, the trigger is not activated. However,
after the trigger condition clears, the trigger is activated
when the condition next occurs.