47
_____________________________________________________________________
3.5 Trigger Functions
______________________________________________________________
Procedure
Screen: CH SETTING / Waveform Display
Flashing cursor
Flashing cursor
3.5.4 Period Trigger
This function sets both the period reference voltage and the
period range, and measures the rise (fall) period of the
reference voltage. When the measured period deviates from
the specified range, triggering occurs.
1. Press the CH SET key to display the desired
CH SETTING screen.
2. Move the flashing cursor to the TRIG position and select
PRIOD using the
keys.
3. Move the flashing cursor to the position shown in the
figure below, and set the lower limit of the trigger level
(period range) using the
keys.
4. Move the flashing cursor to the upper limit, and set the
upper limit of the trigger level (period range) using the
keys.