177
_____________________________________________________________________
11.2 RMS Value
______________________________________________________________
-141 V
3-period
141 V
1/800seconds
Sample the 50 Hz or 60 Hz waveforms at 250
µ
s
and calculate one RMS value item using three-
period data.
Afterwards, the RMS data is calculated in 1/800
second intervals.
11.2 RMS Value
Input data sampled at a fixed 250
µ
s rate is subjected to
800 RMS calculations per second.
One division is equal to 80 samples.
Calculate one RMS value
The maximum and minimum values are specified based on
the RMS-value data (by multiplying 600 RMS-value data
items per second by the time-axis range [s/division]/80),
and the data containing these maximum and minimum
values are defined as one item of sample data. When the
time-axis range is set to 1 min/division, the maximum and
minimum values are specified in the RMS-value data
(800
X
60 [s/division]/80 = 600), and the data containing
these maximum and minimum values are defined as one
item of sample data.