Cleaning and Disinfecting
80
Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Table 1: Approved Cleaners/Disinfectants
*Bleach is not recommended as the primary cleaner/disinfectant.
Remove any disinfectant residue prior to and after the use of bleach
with a new or clean
cloth/wipe soaked in tap water.
When you perform the detailed cleaning steps, please note the following:
•
A microfiber cloth or the Clorox HealthCare® Bleach Germicidal Wipe is recommended as the
wiping cloth
.
•
Always replace the wiping cloth when visibly soiled.
•
Always replace the wiping cloth between steps (spot clean, clean, and disinfect).
•
Always use Personal Protective Equipment (PPE).
•
Adjust the bed position, siderails, headboard, and footboard as needed for ease of cleaning and
disinfection.
Prepare the Bed for Cleaning and Disinfecting
a.
Fully extend the foot section and powered width extensions.
b.
Unplug the bed.
c.
If it is necessary to remove the support surface, go to “Install the Surface and Air Supply Unit”
on page 57, and do the installation steps in reverse order.
STEP 1: Cleaning
a.
As necessary, first remove visible soil from the bed and the support surface using a wiping cloth
soaked with an approved cleaner/disinfectant (see “Table 1: Approved Cleaners/Disinfectants”
on page 80).
•
Give special attention to seams and other areas where soil may accumulate.
•
A soft bristle brush may be used to loosen hardened soil.
•
Use as many wiping cloths as needed to remove the soil.
NOTE:
If desired, the Compella™ support surface cover may be removed and laundered to remove visible
soil. See “Laundry Guidelines” on page 82.
It is important to remove all visible soil from all areas before continuing to remove non-visible soil.
Cleaner/Disinfectant
Recommended for
Routine Cleaning
and Disinfection
Recommended for
Disinfection against
Clostridium Difficile
(C.Diff)
Maintain Wetness
(Disinfection
Contact Time)
Wex-Cide™ Germicidal
Detergent ready-to-use
Yes
No
10 minutes
Virex® II 256
Yes
No
10 minutes
OxyCide® Daily Disinfectant
Cleaner
Yes
Yes
3 minutes
Clorox HealthCare® Bleach
Germicidal Cleaner ready-to-
use
No
*
Yes
5 minutes
Clorox HealthCare® Bleach
Germicidal Wipes
No
*
Yes
3 minutes