Specifications
Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
97
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The P7800 is intended for use in the electromagnetic environment specified below. The customer or the user of the
P7800 should make sure it is used in such an environment.
Immunity Test
EN & IEC 60601-1-2
(4th Ed.)Test Level
Compliance Level
Electromagnetic
Environment—Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
± 8 kV Contact
± 15 kV Air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical Fast
Transient (EFT)
IEC 61000-4-4
± 2 kV for Power Supply
Lines
(100 kHz Repetition
Frequency)
± 2 kV for Power Supply
Lines
(100 kHz Repetition
Frequency)
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV Line(s) to Line(s)
± 2 kV Line(s) to Earth
± 1 kV Line(s) to Line(s)
± 2 kV Line(s) to Earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage Dips
IEC 61000-4-11
(See Note 1)
0% U
T
for 0.5 cycles at
0°, 45°, 90°, 135°, 180°,
225°, 270°, and 315°
(for single-phase mains
equipment)
0% U
T
for one cycle
70% U
T
for 25/50 Hz and
30/60 Hz cycles
0% U
T
for 0.5 cycles at
0°, 45°, 90°, 135°, 180°,
225°, 270°, and 315°
(for single-phase mains
equipment)
0% U
T
for one cycle
70% U
T
for 25/50 Hz and
30/60 Hz cycles
Mains power quality
should be of a typical
commercial or hospital
environment. If the user of
the P7800 requires
continued operation
during power mains
interruption, it is
recommended that the
P7800 be powered from
an uninterruptible power
supply or a
battery.
Voltage Interruptions
IEC 61000-4-11
(See Note 1)
0% U
T
for 250/50 Hz and
300/60 Hz cycles
0% U
T
for 250/50 Hz and
300/60 Hz cycles
Power Frequency (50/60hz)
Magnetic Fields
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a
typical commercial or
hospital environment.
Note 1: U
T
is the AC mains voltage prior to application of the test level.