![Hillrom Compella Instructions For Use Manual Download Page 113](http://html.mh-extra.com/html/hillrom/compella/compella_instructions-for-use-manual_2134337113.webp)
Specifications
Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
99
IMMUNITY to Proximity Fields from Radio Frequency Wireless Communications Equipment
In addition to the Radiated RF IEC 61000-4-3 as shown in the table above, the P7800 has been tested as specified in the
table below.
Test
Frequency
(MHz)
Band
(MHz)
Service
Modulation Maximum
Power (W)
Distance (m)
Immunity Test
Level (V/m)
385
380-390
TETRA 400
Pulse
modulation
18 Hz
1,8
0,3
27
450
430-470
GMRS 460,
FRS460
FM + 5 kHz devi-
ation
1 kHz sine
2
0,3
28
710
704-787
LTE Band 13, 17
Pulse
modulation 217
Hz
0,2
0,3
9
745
780
810
800-960
GSM 800/900
TETRA 800,
iDEN 820,
CDMA 850, LTE
Band 5
Pulse
modulation 18
Hz
2
0,3
28
870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE Band
1, 3, 4, 25;
UMTS
Pulse
modulation 217
Hz
2
0,3
28
1845
1970
2450
2400-2570
Bluetooth
WLAN, 802.11
b/g/n, RFID
Pulse
modulation 217
Hz
2
0,3
28
5240
5100-5800
WLAN
802.11
a/n
Pulse
modulation 217
Hz
0,2
0,3
9
5500
5785