Surface Features and Controls
Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
61
•
Press the highlighted
Max Inflate
option
to stop Max inflate mode immediately.
•
Do nothing—Max Inflate mode ends at
the end of the 1-minute countdown and
the surface returns to the previous mode.
Fowler Boost Mode
When the unit is turned on, it automatically defaults to
Fowler Boost
mode. For
patients in an inclined position,
Fowler Boost
mode increases support in the seat
zone.
NOTE:
When the unit is in Fowler Boost mode, the
Fowler Boost
indicator is on.
To turn Fowler Boost mode on or off, do as follows:
1.
At the Pressure Redistribution screen, press
Mattress Adjust
. The
Mattress Adjustments
screen shows.
2.
Press
Fowler Adjust
. The
Fowler Boost
screen
shows.
3.
Press
On/Off
to turn Fowler Boost mode
On
or
Off
.
4.
If necessary, adjust the setting as follows:
•
To increase the pressure, press
Increase
.
•
To decrease the pressure, press
Decrease
.
•
Press
Default
to return to the standard Fowler Boost setting of
30%
more than the seat
section’s set pressure.
5.
Press
Enter
to return to the previous therapy screen.
Lock or Unlock the Control Panel
1.
At the Pressure Redistribution screen, press
Enter
. The
Patient Setup
screen shows.
2.
Press
Lck Keypd
. The control panel is locked
(or unlocked) and the previous therapy screen
shows.
NOTE:
When the control panel is locked, the
Lockout
indicator shows in the upper right-hand
corner of the screen.