![Hillrom Compella Instructions For Use Manual Download Page 103](http://html.mh-extra.com/html/hillrom/compella/compella_instructions-for-use-manual_2134337103.webp)
Troubleshooting
Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
89
P
OWERED
W
IDTH
E
XTENSIONS
WILL
N
OT
E
XTEND
OR
R
ETRACT
NOTE:
The bed and air supply unit
must
be plugged into AC power for the extensions to operate.
Make sure of these:
•
The bed and the air supply unit are plugged into AC power.
•
The communication cable is connected to the bed and the air supply unit.
•
The air supply unit is turned on.
If the air supply unit is plugged in, but the bed is
not plugged in and/or the communication cable is
not connected, then the
Communication Lost
message will show to allow you to operate the
system in Manual mode until the bed is plugged
into AC power and the communication cable is
connected.
If the system meets the three bulleted conditions
above and the powered width extensions still do
not operate, the air supply unit may have an
internal communication board error. If so, the
Communication Error
message will show.
If you have pressed the
Enter
button, and you need to see the error message or confirm that there is still
an issue, turn the air supply unit off, and then turn it on. If there is still an issue, the
Communication
Error
message will show. If there is an internal communication board error, the system will continue to
operate in the Manual mode until the air supply unit can be serviced or replaced.
T
HE
H
EAD
SECTION
W
ILL
NOT
R
ISE
OR
LOWER
Make sure of these:
•
The Head Up and Down controls are not locked out.
•
The bed has battery charge or is plugged into AC power.
•
The bed CPR release handle is in the fully returned position.
•
The bed width extensions are in the fully extended or fully retracted positions.
3
Side Deflate
Timeout
The system was
left with its
sides deflated
(Manual mode
only).
Yes,
warning
on the
display.
Occurs 30
minutes after
Side Deflate
completes.
Do one of these:
•
Extend Side Deflate.
•
Inflate the sides.
Priority
Condition
Indication
Visual
Warning
Alarm
Condition
Delay
Resolution