1 5 Technical data
55000-00011_002_01 – 1756789 – 2017-09-06
86
15.3
EMC information
EMC notes
Operate the device only with the stated accessories.
The device may only be operated with the accessories stated in the accompanying documents. Operation with
accessories, converters or cables other than those stated in the accompanying documents can lead to increased
EMC emissions or reduced interference immunity of the device and with it to improper use.
Using the device
Use of the device directly adjacent to other devices or stacked with other devices should be avoided, as this can
result in faulty operation. If this form of use is, however, necessary, this device and the other devices should be
monitored to ensure they are working properly.
The lighting system is intended for use in the environment specified below. The customer or user of the device should ensure
that it is operated in one of the environments as described.
Essential characteristics:
The key performance characteristics are the supply of illumination and the limiting of energy in the surgical area.
The recommendations in the instruction manual to preserve BASELINE SAFETY and ESSENTIAL PERFORMANCE PROPERTIES of
the devices during its entire operating life must be observed. See Chapter 11 "Inspections, maintenance and repairs".
WARNING
Measurement of interfering
emissions
Compliance
Comments
RF emissions in accordance with CISPR
11
Group 1
The lighting system uses RF energy exclusively for its
internal FUNCTION. Therefore, its' RF emission is very low
and it is very unlikely that neighbouring devices are
disturbed.
RF emissions in accordance with CISPR 11 Class A
The TruLight lighting system is suitable for use in
establishments other than domestic and those connected
directly to the PUBLIC LOW-VOLTAGE NETWORK that
supplies buildings used for domestic purposes, providing
the following warning is observed:
REMARK
The properties of this device determined by its
EMISSIONS allows its use in industrial fields and in
hospitals (CISPR 11, Class A). If used in a domestic setting
(for which Class B is generally required in accordance with
CISPR 11), this device may cause radio interference. The
user may be required to take adequate measures, such as
conversion or realignment of the device.
Harmonic Emissions as per IEC 61000-3-2 Class A
Harmonic emissions from voltage
fluctuations and flicker emissions as per
IEC 61000-3-3
In compliance