2 Safety instructions
55000-00011_002_01 – 1756789 – 2017-09-06
done by an authorised hospital technician or a person with equivalent
qualification.
Deinstallation for service purposes
The spring arm may bounce up
When the light head/flat screen is removed before moving
the spring arm to the top-most limit stop position, the spring
arm will shoot up and can cause severe injuries:
• The light head/flat screen may therefore only be de-installed by
Trumpf Medical Customer Service.
Commissioning
Initial commissioning prior to use
The light system must be handed to the user in a tested state after initial
operation before it can be used in routine medical procedures.
• The initial operation includes functional and safety checks on the
entire lighting system.
• Handover must be documented by a handover declaration.
WARNING
CAUTION