Operation
8.1
Working rules
Compliance with safety instructions
The safety instructions must be read and adhered to when using the lighting system
to ensure that it is safely handled and to prevent harm to the patient.
8.2
Preparatory measures
Measures that should be considered before using the lighting system.
Checking the lighting system
Function test / visual inspection
A function test and a visual inspection should be carried out before use, or at least
once per week.
Contamination and infection hazard for patients
Loose or damaged parts may fall into wounds. To ensure the safety of
patients, check the components of the lighting system for the following
points before each use:
• Loose parts on light head,
• visible damage, in particular on the cover plates of the light head and
the sterile handle,
• secure mounting of the sterilisable handle.
---------------------------------------------------------------------------------------
Electric shock
There is a risk of electric shock in the event of contact with
damaged electrical components of the mobile stand version:
• Do not connect the lighting system to the mains in the event of
damaged plug connectors or a damaged mains cable.
• Take the device out of service and label it as DEFECTIVE.
---------------------------------------------------------------------------------------
LED failure
After failure of the tenth LED, the light head does not achieve the
specified light intensity.
• Take the lighting system out of service.
---------------------------------------------------------------------------------------
Decommissioning
In the event of functional defects or damage which impair the
operational safety of the lighting system, the device must be immediately
taken out of service:
• Disconnect the lighting system by using the master switch installed in
the building, or remove the safety connector.
• Secure the master switch or the safety connector against accidental
switch-on/plugging in.
• Label the lighting system as DEFECTIVE!
• Inform Trumpf Medical Customer Service.