6 Commissioning
55000-00011_002_01 – 1756789 – 2017-09-06
45
Figure 21
6.3
Positioning the mobile stand version
The mobile stand version has a high tilting stability. The
following, basic precautions must nevertheless be considered
when positioning the mobile stand version.
Tilting of the stand
The tilting stability of the mobile stand version can
be at risk due to objects lying on the floor, uneven
floors or the mains cable:
• Always move the mobile stand to the place of
operation with the light head pointing
forwards.
• Do not drive over objects lying on the floor,
uneven parts of the floor or the mains cable.
• Pull the mains plug out of the socket and roll
up the mains cable ④ at the power supply
housing ⑤ to move the stand.
------------------------------------------------------------
The pedestal might lt when the castors ⑦ are
locked and excessive force is exerted onto the
spring arm ② or light head ①:
• Avoid strong leverage onto the spring arm or
the light head.
• Do not add additional loads to the spring arm.
Locking the stand base
The two front castors ⑦ can be locked to
immobilise the stand base ⑥.
• Press the brakes ⑧ of both front castors
downwards to lock them.
• Press the brakes ⑧ of both front castors
upwards to unlock them.
Positioning the stand
1. Release the brakes on the two front castors.
2. Grasp the stand by the stand rod ③ and move it with
the light head ① in the direc on of travel to the place of
use.
3. Posi on the light head ① by using the sterile handle or
the outer non-sterile outer handles at a working distance
of 70–150 cm from the wound area.
4. Connect the mobile stand version to the mains.
5. Align the light head ① with the wound area using the
non-sterile handle.