2 Safety instructions
55000-00011_002_01 – 1756789 – 2017-09-06
Contact with laser beams
Damage to vision
Direct contact with the laser beam may cause damage to
vision:
• Do not look directly into the laser beam
• Protect the patient's eyes:
The patient's eyes must be closed or protected as necessary (e.g.
with safety goggles with an optical density of at least 2 or designed
according to protection level 6 EN169).
• Use of operating methods or procedures other than those stated in
these instructions for use may result in dangerous radiation effects
due to the laser.
Light intensity of the light head
LED failure
After failure of the tenth LED, the light head does not achieve the
specified light intensity.
• Take the lighting system out of service.
• Inform Trumpf Medical Customer Service. Exchanging a light head or
repairs to the lighting system may only be performed by Trumpf
Medial Customer Service or by service staff trained and authorised
by Trumpf Medical.
Additional loads
Crashing of the light system
Do not place any additional weight on the light system
Swivel movement of the light head
Risk of injury due to uncontrolled swivel movement
Uncontrolled movement of the spring arm may result when
the spring force of the spring arm is not correctly adjusted.
Risk of jamming
When rotating the light head, the distance between the
cardan joint and the light head changes:
• Do not insert your fingers between the cardan joint and the light
head when swivelling the light head.
WARNING
CAUTION
WARNING
WARNING