1 5 Technical data
55000-00011_002_01 – 1756789 – 2017-09-06
83
TruLight 5000 electrical data
TruLight 3000 lighting system technical data
* = at a distance of 0.9 m
Electrical data
TruLight 5300
TruLight 53x0
TruLight 5500
TruLight 55x0
Mains adapter supply voltage
100 - 240 VAC
50/60 Hz
100 - 240 VAC
50/60 Hz
100 - 240 VAC
50/60 Hz
100 - 240 VAC
50/60 Hz
Supply voltage at the DC-DC converter
19 - 36 V DC
21.6–26.4 V AC
19 - 36 V DC
21.6–26.4 V AC
19 - 36 V DC
21.6–26.4 V AC
19 - 36 V DC
21.6–26.4 V AC
Light head power uptake
(160,000 lux at a distance of 1.0 m)
65 VA
Max. power uptake of total system
110 VA
140 VA
120 VA
160 VA
Internal fuse (mobile stand version only)
2 x T10 A
Voltage at fixed point on ceiling
48 V
Average Service Life of Light Source (LED)
> 60,000 hrs
Classification according to MPG
1
1
1
1
Product designation
TruLight 33x0
TruLight 35x0
Light intensity at 1.0 m
140,000 lux
160,000 lux
Dimmable from/to
< 10% Endo;
40%–100%
< 10% Endo;
40%–100%
Variable light field size due to changes in distance
17 - 25 cm
17 - 25 cm
Light field diameter (d10) at 1.0 m
18 cm
18 cm
Light field diameter (d50) at 1.0 m
9.4 cm
9.5 cm
d50/d10 ratio
0.52
0.53
Radiant power (W/m)*:
587
616
Residual light intensity with 1 switch
81,200 lux
58%
108,800 lux
68%
Residual light intensity with 2 switches
61,600 lux
44%
75,200 lux
47%
Residual light intensity with tube
140,000 lux
100%
155,200 lux
97%
Residual light intensity with tube and 1 switch
81,200 lux
58%
104,000 lux
65%
Residual light intensity with tube and 2 switches
61,600 lux
44%
72,000 lux
45%
Illumination depth (L1 + L2) at 20%
Ec / EN ISO 60601-2-41 2nd Edition
83 cm
75 cm
Illumination depth (L1 + L2) at 60%
Ec / EN ISO 60601-2-41 3rd Edition
44 cm
41 cm
Colour rendering index Ra
94
93
Colour temperature
4.500 K