Cleaning and disinfection
Regular cleaning and disinfection with suitable cleaning or disinfection agents is
necessary for the safe use of the surgical light.
Electric shock
Touching live components may result in an electric shock.
• Disconnect the device before cleaning and disinfection:
• Ensure that no cleaning or disinfectant fluids penetrate into the
device or into the support arm system.
• Do not place any objects in the equipment openings.
Ceiling-mounted and wall-mounted versions
• Switch off the operating theatre master switch.
→ The lighting system is disconnected from the power supply.
Mobile stand version
• Switch off the power supply unit, position the switch in Position
0
(see Chapter
6.3.1, page 46), pull the mains cable out of the safety socket.
→ The mobile lighting system is disconnected.
Improperly used cleaning agents or disinfectants can
pose a risk for patients or damage products
If the information and instructions contained in this chapter
are not observed or complied with, this may result in a risk of
contamination or infection for the patient or damage to the product.
Furthermore, it would render any claim for damages void!
• Dispense cleaning agents and disinfectants such that no liquid can
enter through joints or openings of the surgical lamp or parts of the
support arm system.
• Use the surface disinfectant only at the concentration specified by
the manufacturer.
• Only use disinfectants approved by the manufacturer for use with
the following materials:
Polycarbonate (PC), polyamide (PA), acrylonitrile butadiene styrene
copolymer (ABS), polystyrene (PS), polyurethane (PUR),
polyphenylene sulphone (PPSU), polybutylene terephthalate (PBT)
and silicones.
• In the event of an excessive coating of surface disinfectant forming,
carry out a thorough cleaning.
• Due to the risk of surface damage:
-
Do not use sharp, pointed or abrasive objects,
-
Do not use abrasive substances or agents which can remove material,
-
Do not use solvents, benzene, paint thinners, alkaline cleaning agents or
cleaning agents containing acids or aldehydes,
-
Do not use agents with glycol derivatives, phenols, phenol derivatives or
quaternary compounds,
-
To prevent paint or corrosion damage, only use agents that do not contain
chlorides or halogenides.
WARNING
WARNING