![Hill-Rom 4038110 Instruction Manual Download Page 34](http://html1.mh-extra.com/html/hill-rom/4038110/4038110_instruction-manual_2134156034.webp)
5 Description of devices and functions
55000-00011_002_01 – 1756789 – 2017-09-06
34
Figure 10
5.4.3
Controlling the light intensity
Non-sterile setting of the light intensity
The light intensity is set at the
Light
③
control panel.
•
A:
for TruLight 3000 ① at the control
panel ② for the light head or the
op onal wall-mounted control panel ④,
or
•
B:
for TruLight 5000 ⑦ on the control
panel ⑤ of the control unit ⑥, or on the
op onal wall-mounted control panel ⑧.
The intensity of illumination of the light head
can be adjusted in 7 stages:
– < 10 % (Endo)
– 40 %–100 %
Reducing the light intensity:
• Press key ⑨ on the control panel.
→ The LED for the currently set light
intensity ⑪ lights up.
Increasing the light intensity:
• Press key ⑫ on the control panel.
→ The LED for the currently set light
intensity ⑪ lights up.
Endo-dimming
The light intensity level Endo-dimming ⑩ is
intended as the light intensity for endoscopic
operations.
Activating Endo-dimming:
• Press key ⑨ on the control panel un l
the Endo LED lights up.
→ The light intensity is reduced to a value <
10 % of the maximum light intensity.
Deactivating Endo-dimming:
• Press key ⑫ on the control panel un l
the Endo LED goes out.
The light intensity is set to a value in the range
of 40 %–100 % (indicated by the