6 Commissioning
55000-00011_002_01 – 1756789 – 2017-09-06
44
Figure 20
Wall-mounted version single light
C:
Surgical light as single light with a light head on an AC 2000
or AC 2000 LCH (Low Ceiling Height) type spring arm.
• The best possible mobility of the support arms is a
"V-type arrangement”.
Working distance
Position the light head with the sterile handle or the outer
non-sterile outer handles at a working distance of
70–150 cm from the wound area.
Adaptive Light Control plus (ALC plus), optional
In order to use the functionality of Adaptive Light Control plus
effectively, the integrated distance measurement must make
its measurement within the wound area. The sterile handle of
the light head must therefore be adjusted so that the laser
beam falls inside the wound area.
Damage to vision
Direct contact with the laser beam may
cause damage to vision:
• Do not look directly into the laser beam
• Protect the patient's eyes:
The patient's eyes must be closed or protected
as necessary (e.g. with safety goggles with an
optical density of at least 2 or designed
according to protection level 6 EN169).
• Use of operating methods or procedures other
than those stated in these instructions for use
may result in dangerous radiation effects due
to the laser.
Switch the power supply to Standby
• Switch on the operating theatre master switch.
• Switch on light head (see chapter Operation).