55000-00011_002_01 – 1756789 – 2017-09-06
66
11
Inspections, maintenance and repairs
All devices are subject to wear and tear over time. The safety and function of the
device must therefore be inspected after regular inspection and maintenance
intervals. Repeat inspections must be carried out according to the specific national
regulations. Trumpf Medical recommends taking out a service contract.
Technical Customer Service
Telephone: +49 3671 586-41911
Fax: +49 3671 586-41175
[email protected]
11.1
Inspections during operation
Weekly inspection
A function test and visual inspection must be performed before each use of the
lighting system in running medical operation, however at least once per week.
Contamination and infection hazard for patients
Loose or damaged parts may fall into wounds. To ensure the safety of
patients, check the components of the lighting system for the following
points before each use:
• loose parts in the light heads,
• visible damage, in particular on the cover plate of the light head and
the sterilisable handle,
• secure mounting of the sterilisable handle.
---------------------------------------------------------------------------------------
Defective devices
Defective devices or functional units must be clearly labelled immediately
and taken out of operation.
Contact Trumpf Medical Customer Service or an authorised service
partner in the event of damage or faults.
11.2
Annual visual inspection
Annual visual inspection
A visual inspection of the hardware of the entire lighting system must be carried out
annually by the operator's qualified technicians. In the visual inspection, in particular
the following parts and components of the device must be examined for changes to
the material:
• Deformation of components of the light head and the support system,
• Paint damage on the entire support arm system and on the light heads,
• missing plastic components and small parts, e.g. covers, plugs, etc.
• cracking and brittle plastic parts,
• legibility of type labels.
Damaged or deformed components must be replaced.
WARNING