8 Operation
55000-00011_002_01 – 1756789 – 2017-09-06
58
Figure 31
Fine positioning Adaptive Light Control Plus (ALC plus)
To use Adaptive Light Control plus (ALC plus), point the light
head precisely at the wound area ④ using the sterile handle
②.
Activating the ALC plus function:
1. Press the key ⑥ Auto.
→ The LED ⑦ lights up and indicates that ALC plus is
activated.
2. Briefly readjust the position of the light head:
→ The laser beam is switched on automatically and
measures the distance of the light head from the wound
area ④ again.
→ The light intensity is immediately and precisely adapted
to the working distance on the basis of the measurement
data.
Deactivate the ALC plus function:
3. Press the key ⑥ Auto.
→ The LED ⑦ goes out and indicates that ALC plus is
deactivated.
→ Only the functions of Adaptive Light Control (ALC) are
now available.
Manual/auto switchover
Switchover between automatic distance detection
and manual adjustment of the distance is
performed by pressing the keys Auto ⑥ or Manual
⑧.