9 Cleaning and disinfection
55000-00011_002_01 – 1756789 – 2017-09-06
Cleaning procedure
1. Disconnect the power supply from the mains.
2. Allow the light head to cool down. Only clean or disinfect the light head when it
is cold.
3. Moisten a cloth with a drop of cleaning or disinfecting agent. Clean the surgical
light with the damp but not wet cloth.
Risk of fire or explosion by disinfectants
Production of gases, fumes or mists when using disinfectants
may create a combustible or explosive atmosphere.
• Do not use highly flammable disinfectants.
• Do not disinfect large areas.
• Allow hot surfaces to cool down before disinfection.
• Where possible, completely isolate the room’s electrical systems or
ensure that no switching processes – especially automatic processes
– are initiated or run whilst disinfection is in progress.
• After wipe-over disinfection, wait until the disinfectant has
completely dried.
• Ensure the room is adequately ventilated.
Comply with national guidelines.
The operator must observe the requirements of the responsible national
hygiene and disinfection board.
9.1.3
Recommended disinfectants
Trumpf Medical recommends the following disinfectants for manual application:
• The list of approved disinfectants is continually updated and is given in the ‘OP
lights instruction manual as a supplement' which can be downloaded in the OIS.
WARNING
NOTE
Manufacturer
Product designation
B. Braun Melsungen AG
Meliseptol
Various
70 % 2-propanol alcohol
Schülke & Mayr GmbH
Perform 0.5 %
Bode Chemie GmbH & Co. KG
Dismozon pure 0.75 %
Clorox Healthcare
Hydrogen peroxide cleaner disinfectant wipes
Kesla Pharma
Wofasteril 0.5 %