20
2 Safety instructions
55000-00011_002_01 – 1756789 – 2017-09-06
• Only position the light head by using the sterile handle or the non-
sterile handles (outer handles).
Cleaning and disinfection
Improperly used cleaning agents or disinfectants can pose a risk
for patients or damage products
If the following information and instructions are not observed or
complied with, this may result in a risk of contamination or infection for
the patient or damage to the product. Furthermore, it would render any
claim for damages void!
• Use the wipe-over method only for disinfection.
• For cleaning and disinfection, only wipe with a damp but not wet
cloth.
• Dispense cleaning agents and disinfectants so that no liquid can
enter through joints or openings of the surgical light or parts of the
support arm system.
• Use the surface disinfectant only at the concentration specified by
the manufacturer.
• Only use disinfectants approved by the manufacturer for use with
the following materials:
Polycarbonate (PC), polyamide (PA), acrylonitrile butadiene styrene
copolymer (ABS), polystyrene (PS), polyurethane (PUR),
polyphenylene sulphone (PPSU), polybutylene terephthalate (PBT)
and silicones.
• In the event of an increased layer formation of surface disinfectant,
thorough cleaning must be performed.
• Due to the risk of surface damage:
–
Do not use sharp, pointed or abrasive objects,
–
Do not use abrasive substances or agents which can remove material,
–
Do not use solvents, benzene, paint thinners, alkaline cleaning agents or
cleaning agents containing acids or aldehydes,
–
Do not use agents with glycol derivatives, phenols, phenol derivatives or
quaternary compounds,
–
To prevent paint or corrosion damage, only use agents that do not contain
chlorides or halogenides.
• It is essential that the hygiene instructions of the operator are
observed.
For more detailed information about cleaning and disinfection, see Chapter 9, page 60
Adjustments
Adjusting the device
The manufacturer guarantees the safety and proper working of the
device only under the condition that the work of adjustment has been
WARNING
CAUTION