6 Commissioning
55000-00011_002_01 – 1756789 – 2017-09-06
46
Figure 22
6.3.1
Connecting the mobile stand version to the
mains
The power supply of the mobile stand version is established
by a mains cable ⑥ with an IEC connector ③ and a safety
connector ⑤.
Electric shock
There is a risk of electric shock in the
event of contact with damaged
electrical components:
• Do not connect the lighting system to the
mains in case of defective plug connectors
③/⑤ or in case of damaged mains cables
⑥.
• Label the device as DEFECTIVE and contact
Trumpf Medical Customer Service.
------------------------------------------------------------
Earthing of the mains socket
In case of an electrical short circuit, earthing
reduces the danger of electric shock:
• The mobile stand version (Protection Class I)
may only be connected to a properly earthed
safety socket.
Connect the power supply to the mains
1. Check whether the mains voltage corresponds to the
information on the name plate. Ask the local energy
supply company or a specialist electrical company in
cases of doubt.
2. Lay the mains cable ⑥ so that there is no danger of
tripping and so that there is no strain on the cable.
3. Plug the IEC power connector plug connector ③ into the
socket ④ on the underside of the power supply housing
①.
4. Insert the mains plug ⑥ into a correctly installed and
earthed three-pin plug socket.
Switch the power supply to Standby
• Switch on the power supply unit with the ON/OFF switch
② on the power supply housing.
• Switch on light head (see chapter Operation).