Troubleshooting
179
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
P/N 130859-US, Manual Revision: AA
143
Device Cover Open
Explanation:
The device cover is open. The device cover must be closed and locked for the procedure to
continue.
Corrective Action:
Close the device cover.
If the problem persists, service is required.
144
Pump Communication Failure
Explanation:
An issue has been detected with the pump.
Corrective Action:
Restart the device.
If the problem persists, service is required.
145
Pump Failure
Explanation:
An issue has been detected with the pump.
Corrective Action:
Restart the device.
If the problem persists, service is required.
146
Unable to Prime
Explanation:
No fluid has been detected while attempting to prime the yellow line.
Corrective Action:
1. Check the yellow line for kinks and occlusions.
2. Ensure the clamps on the yellow line are open and saline is connected.
3. Check the tubing placement in the air detector and pump.
4. Touch
Continue
to continue.
Table 22, Event Messages