Remove and Reinstall Effluent Line Tubing
Explanation:
The effluent line sensor is looking for tubing and does not detect it. This can occur due to
calibration problems or if the tubing is not fully seated in the effluent line sensor.
Corrective Action:
1. Ensure the tubing is fully seated in the effluent line sensor.
2. If the tubing is fully seated, remove it, clean the effluent line sensor, and reinstall the tubing.
3. Close the device cover.
248
Duplicate Settings Group Name
Explanation:
There cannot be two settings groups with the same name.
Corrective Action:
Add a group again and enter a name that does not already exist.
250
Invalid Settings Group Name
Explanation:
A name cannot be only spaces. Characters must be used.
Corrective Action:
Add a group again and enter a name that contains valid characters.
252
Stop Key Failure
Explanation:
An issue has been detected with the
Stop
key.
Corrective Action:
Restart the device.
If the problem persists, service is required.