IEC/EN 60601-1-2 Standard Requirements
221
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
P/N 130859-US, Manual Revision: AA
Table 27, Test specifications for enclosure port immunity to RF wireless communications
equipment
The system is intended for use in the electromagnetic environment specified below. The customer or
user of the system should ensure that it is used in such an environment.
Test
Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
Power
(W)
Distance
(m)
Immunity
Test Level
(V/m)
385
380-390
TETRA 400
Pulse
modulation
b)
18 Hz
1.8
0.3
27
450
430-470
GMRS 460,
FRS 460
FM
c)
+5 kHz deviation
1 kHz sine
2
0.3
28
710
704-787
LTE Band 13,
17
Pulse
modulation
217 Hz
0.2
0.3
9
745
780
810
800-960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
1845
1970
2450
2400-2570
Bluetooth;
WLAN,
802.11 b.g.n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0.3
28
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used, because while it does not represent actual modulation, it would be worst case.