Wash Line Priming
(On/Off)
Determines if the device omits the Wash Prime process at the beginning of the
first Fill phase to reduce the duration of the procedure.
On:
The device performs the Wash Prime process at the beginning of the
first Fill phase.
Off:
The device performs the Wash Prime process at the beginning of the
first Wash phase.
Recentrifuge Delay
(On/Off)
Determines if device performs the recentrifuge delay. For more information,
see
“Recentrifuge Delay” on page 91
Note: This must be set separately for Cell Salvage and Manual Mode.
On:
The device performs the recentrifuge delay. The user can override
the recentrifuge delay after 15 seconds by touching
(Play
).
Off:
The device does not perform the recentrifuge delay (the device does
not wait for the centrifuge to ramp up before starting the pump).
Attention:
If the delay is turned off, be aware of the possibility of red cell
spillage caused by pumping fluid into a bowl of red blood cells with poor
or no separation.
Parameters
Cell Salvage
Fill Start Volume:
The approximate volume of fluid in the reservoir that
transitions the device from Standby to Fill when the bowl is empty.
Fill Resume Volume:
The approximate volume of fluid in the reservoir
that transitions the device from Standby to Fill when the bowl already
contains some fluid.
Fill Pump Rate:
The approximate rate at which the pump turns while
filling the bowl with fluid.
Minimum Wash Volume:
The minimum volume of wash solution to be
used to wash one bowl of fluid.
Wash Pump Rate:
The approximate rate at which wash solution enters
the bowl.
Smart Empty:
Determines if the device slows the pump speed at a
decremental rate to empty fluid from the bowl or if the pump speed
remains constant. This setting is not available for the 70 mL bowl.
On:
The pump empties fluid from the bowl at a preprogrammed
decremental speed determined by the parameter settings.
Off:
The pump empties fluid at a constant speed, which the user can
adjust via the
+/-
pads from 25 mL/min to 1000 mL/min (default is 100
mL/min).