302
Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
• CLOSE the white air line clamp.
• CLOSE the blue PRP line clamp.
• OPEN the yellow PPP line clamp.
2. Agitate the blood bag.
3. Touch
Continue
to fill the bowl.
303
Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
• CLOSE the white air line clamp.
• CLOSE the blue PRP line clamp.
• OPEN the yellow PPP line clamp.
2. Touch
Continue
to collect PPP.
304
Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
• CLOSE the white air line clamp.
• OPEN the blue PRP line clamp.
• CLOSE the yellow PPP line clamp.
2. Touch
Continue
to collect PRP.
Table 22, Event Messages
ID #
Event Message Text