136
Chapter 7
P/N 130859-US, Manual Revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
Making the
Quick Transfer
and Manual
Mode Settings
Available
Two additional modes can be made available on the Cell Saver Elite+ device:
Quick Transfer Mode enables the device to quickly move
unwashed
shed
blood from the reservoir to the reinfusion bag and bypass the wash phase
completely when time is of the essence.
Manual Mode enables the device to be operated manually by the user.
Note: These modes are only available when using the Cell Salvage protocol.
Note: These modes can only be made available in a settings group. To create
a new settings group, see
“Creating a New Settings Group” on page 125
Note: Contact your local Haemonetics representative for access to training and
the password to make this feature available
To make the Quick Transfer or Manual Mode setting available:
1. Touch
(
Menu
).
2. Select
System
from the drop-down list. The
System
screen appears.
3. Touch
Unlock
. The
Enter Access Code
screen appears.
4. Enter the access code and touch
(
Accept
).
5. On the
System
screen, touch
Options
. The Option screen appears.
6. Touch either
Quick Transfer
or
Manual Mode
.
7. Touch
(
Up)
or
(
Down
)
to change the setting to
Yes
. If desired,
select the other mode and change its setting to
Yes
.
8. Touch
Done
.
The enabled functionality is now available for use in the settings group. To
utilize Quick Transfer, see