General Operation: Cell Salvage
81
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
P/N 130859-US, Manual Revision: AA
Figure 38, Inserting the bowl into the centrifuge chuck
4.
70 mL bowl only:
Ensure that the blue indicator lines inside the chuck
adaptor are visible.
5. Position the header arm around the top of the bowl.
6. Turn the latch on the header arm clockwise until it locks into place.
7. Spin the bowl to ensure it spins freely.
Figure 39, Closing and locking the header arm
Note: A click will be heard when the locking mechanism is completely secured.
Alert: Verify that the outlet port and effluent tubing are free of any
restrictions prior to initiating a procedure. If the outlet port is
inadvertently clamped off, the bowl rotary seal may become
compromised. See
“Avoiding Flow Restrictions” on page 65
for more
information.
Installing the Effluent Line Sensor Tubing
To install the effluent line tubing:
1. Floss the effluent tubing into the effluent line sensor groove.
2. Ensure that the tubing is completely installed in the effluent line sensor.
1.
2.
3.
4.
1. Header arm
2. Tubing in effluent line
sensor
3. Header arm latch
4. Bowl in centrifuge chuck