Note: The device should not be used near active high-frequency surgical
equipment or in the RF shielded room of MRI equipment, where the intensity
of electromagnetic disturbances is high.
Electrical
Specifications
The electrical specifications for operating the Cell Saver Elite+ device are as
follows.
Attention:
The Cell Saver Elite+ device meets the requirements of the IEC/EN
60601-1-2 Standard, Electromagnetic compatibility (EMC).
Additional IEC/EN compliance information is available in
.
Note: The power source used must be properly grounded.
*In accordance with IEC/EN 60601-1standard, medical electrical equipment, general
requirements for safety.
Table 9, Environmental Specifications
Conditions
Values
Ambient operating temperature 10 °C to 27 °C (50 °F to 80.6 °F)
Storage/transportation
temperature
-20 °C to 50 °C (-4 °F to 122 °F)
Operating humidity level
8 to 80% R.H., non-condensing above 0°C
Atmospheric pressure range
< 2438 meters (8000 ft.)
Pollution degree
Pollution degree 2
Table 10, Electrical Input Power
Rated Voltage
Rated Current
Fuse
Frequency
100–120 V
3.0 A
T3.15A 250V
50/60 Hz
200–240 V
1.5 A
T3.15A 250V
50/60 Hz
Table 11, Enclosure/Chassis Leakage Current Specifications*
Condition
Polarity
Ground
Max Value
Normal
Normal
Normal
100 µ
Α
Reverse
Normal
100 µ
Α
Single Fault
Reverse
Open
500 µ
Α
Normal
Open
500 µ
Α