Bowl Optics Failure
Explanation:
An issue has been detected with the bowl optics. Processing has paused. Processing can
continue, but the user must manually start the Wash phase.
Corrective Action:
To proceed manually:
1. Touch
Continue
to continue.
2. Touch
Wash
to start Wash when there is adequate fluid in the bowl.
NOTE:
The bowl is full when the red cell/supernatant interface reaches a point approximately
1/4 inch over the shoulder of the bowl. The hematocrit of the product may be reduced if the Wash
phase is started before the bowl is full.
OR
To end the procedure:
1. Touch
End Procedure
Service is required to fully resolve this issue.
126
Bowl Size Mismatch
Explanation:
A Latham bowl was detected, but a 70 mL bowl was selected.
Corrective Action:
If a 70 mL bowl is installed:
1. Check the bowl and chuck adaptor for proper installation in the centrifuge.
2. Clean the optics.
3. Touch
Keep Bowl Size
.
OR
If a Latham bowl is installed:
1. Touch
Change Bowl Size
.
2. Select the new bowl size or scan the processing set barcode.
Table 22, Event Messages
ID #
Event Message Text