118
Chapter 6
P/N 130859-US, Manual Revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
Changing to a
Cell Salvage
Procedure
You can also choose to end the Sequestration procedure early and change to
a Cell Salvage procedure. To end Sequestration early and change to Cell
Salvage:
1. Touch
(
Menu
) and select
Cell Salvage
from the drop-down list. The
device displays a confirmation message.
2. Touch
Continue
to continue to Cell Salvage. The device displays a
prompt:
Figure 62, Changing to Cell Salvage
3. Following the instructions on the screen, change to a Cell Salvage
processing set.
a. Disconnect the collection bag harness from the effluent tubing.
b. Cap the collection bag line to prevent drips.
c. Connect the effluent tubing to the waste bag.
d. Close the blood bag line clamps.
e. Remove the upper portion of the blood bag harness, using the twist-
lock connector, and cap the port using the attached caps.
f. Open the saline bag clamp.
g. Open the reservoir drain port clamp.
h. Ensure that the slide clamp on the blood bag harness is open.
i. Touch
Cell Salvage
to continue to Cell Salvage.