Troubleshooting
211
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
P/N 130859-US, Manual Revision: AA
311
Operator Action Required
Explanation:
The processing set must be prepared before continuing to Cell Salvage.
Corrective Action:
1. Remove the PPP, PRP, and air bag connection and reconnect the waste bag.
2. Close the blood bag clamps and open the reservoir clamp.
3. Open the saline bag clamp.
4. Touch
Cell Salvage
to continue to Cell Salvage.
OR
1. Touch
Sequestration
to return to Sequestration.
312
Confirmation Required
Explanation:
The Cell Salvage menu option was selected. This will stop Sequestration. Confirmation is
required to proceed.
Corrective Action:
Touch
Cancel
to continue Sequestration.
OR
Touch
Continue
to continue and prepare for Cell Salvage.
OR
Touch
End Procedure
to end the current procedure.
400
System Fault
Explanation:
An issue has been detected with the system.
Corrective Action:
Restart the device.
If the problem persists, service is required.
Table 22, Event Messages
ID #
Event Message Text