220
Appendix A
P/N 130859-US, Manual Revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
Table 26, Recommended separation distance between portable RF communications equipment and
the Cell Saver Elite+ device
The system is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or user of the system can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the system as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
Watts (W)
Separation distance according to frequency of transmitter - meters (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1.0
1.2
1.2
2.3
10
3.7
3.7
7.4
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz, the separation for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
d
3.5
V
1
-------
P
=
d
3.5
E
1
-------
P
=
d
7
E
1
------
P
=