Protocol Settings
135
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
P/N 130859-US, Manual Revision: AA
Empty Pump Rate:
The approximate rate at which the pump turns while
emptying fluid from the bowl. For the 125 mL and 225 mL bowls, this
setting is only adjustable if Smart Empty is set to “Off.” For the 70 mL
bowl, this setting is always available.
Emergency Empty Pump Rate:
The approximate rate at which the
pump turns while emptying fluid from the bowl in Emergency Mode. If
Smart Empty is set to “Off”, this setting can be adjusted by the user via
the
+/-
pads. This setting is not available for the 70 mL bowl.
Emergency Fill Pump Rate:
The approximate rate at which the pump
turns while filling the bowl with fluid in Emergency Mode. This setting is
not available for the 70 mL bowl.
Emergency Wash Pump Rate:
The approximate rate at which wash
solution enters the bowl while in Emergency Mode. This setting is not
available for the 70 mL bowl.
Sequestration
Fill Pump Rate:
The approximate rate at which the pump turns while
filling the bowl with fluid and collecting PPP.
Fill Centrifuge Speed:
The approximate rate at which the centrifuge
turns while the bowl is filling with whole blood and during the PPP
collection phase.
PRP Collection Pump Rate:
The approximate rate at which the pump
turns while collecting fluid during the PRP collection phase.
PRP Collection Centrifuge Speed:
The approximate rate at which the
centrifuge turns during the PRP collection phase.
PRP Extend Volume:
The volume in milliliters to pump after the effluent
line sensor detects RBCs during the PRP collection phase. Once this
volume is pumped, the PRP collection is complete.
PRP Milk:
Determines if the process of starting and stopping the PRP
collection phase is desired as determined by the programmed
parameters “Milk Vol” and “Milk Delay.”
PRP Milk Volume:
The volume to pump during the PRP collection until
the pump stops and the Milk delay time starts. This setting is only
adjustable if PRP Milk is set to “Yes.”
PRP Milk Delay:
The time the pump will be stopped during the PRP
collection phase before starting to pump the programmed Milk Vol
parameter. This setting is only adjustable if PRP Milk is set to “Yes.”
PPP Return Plasma:
Determines which clamps to open and close at the
beginning of the Empty phase and to account for volume returned. If you
select “Yes,” the PPP is returned to the bowl and then to the RBC bag
along with the RBCs.