Reservoir Load Cell Calibration Failure
Explanation:
An issue has been detected with the reservoir sensor calibration during the power-on self-test
(POST).
Corrective Action:
Please service the device before continuing use.
186
Waste Bag Load Cell Calibration Failure
Explanation:
An issue has been detected with the waste bag sensor calibration during the power-on self-test
(POST).
Corrective Action:
Please service the device before continuing use.
187
Manifold Pressure Sensor Calibration Failure
Explanation:
An issue has been detected with the manifold pressure sensor calibration during the power-on
self-test (POST).
Corrective Action:
Please service the device before continuing use.
189
Header Arm is Unlatched
Explanation:
The device detected that the header arm is unlatched. The centrifuge must stop before the cover
will unlock and corrective action can be taken.
Corrective Action:
1. Ensure that the bowl is seated properly.
2. Close and latch the centrifuge arm.
3. Close the device cover.
If the problem persists, service is required.