51
4.
Pour a 1413 µS/cm
conductivity standard to
the fill line on the sensor
cap.
5.
Put the sensor fully into
the sensor cap. The tester
will read the standard
value.
Air bubbles under the
probe tip when submerged
can cause slow
stabilization or error in
measurement. Shake the
tester from side to side to
remove air bubbles.
6.
Push the
Calibration/Settings key
to start the calibration.
7.
Wait for the value to
stabilize, then push the
Calibration/Settings key to
keep the standard value.
8.
The display shows
“END” when the calibration
is complete.
9.
To measure a second
standard (147 µS/cm), do
step 2, then steps 4
through 7 again.