87
9.
Set the 1.0 - 5.0 mL
pipette to 2.0 mL.
10.
Put the tip of the pipette
into the sample and slowly
pull in 2.0 mL.
11.
Slowly dispense the
sample into the test tube in
position A9.
12.
Set the 1.0 - 5.0 mL
pipette to 1.5 mL.
13.
Put the tip of the pipette
into the sample and slowly
pull in 1.5 mL.
14.
Slowly dispense the
sample into the test tube in
position A8.
15.
Do steps
13
and
14
againto put 1.5 mL of the
sample into the test tube in
position A10.
Figure 7 on page 89
shows the contents of the
test tubes in Row A after
this step is completed.
16.
Pull the solution in A9
into the pipette 2 to 3 times
to mix the sample dilution.
17.
Start making the
sample dilution series in
Row A:
Pull in 1.5 mL of solution
from A9 and put it into A7
using the pipette.
Pull the solution in A7 into
the pipette 2 to 3 times to
mix the sample dilution.