37
5.3 EZ Arsenic, 0–4000 ppb (0, 35, 75, 175, 1500, 4000)
5.4 Interferences
1.
Insert a test strip into the
cap so the pad completely
covers the small opening.
Close the flap and press to
secure.
2.
Fill the square
measuring vial to the top
with sample (9.6 mL). Pour
the sample into the reaction
bottle.
3.
Add one Reagent #1
and one Reagent #2
powder pillow to the
sample.
See
Interferences on
page 37
.
4.
Immediately attach the
cap to the reaction bottle.
Swirl continuously for 60
seconds.
Do not shake or invert or
allow sample to get on
the strip.
Wait 20 minutes.
Swirl twice during the
reaction period.
5.
Remove the test strip
and immediately compare
the developed color to the
chart on the test strip bottle
(0–4000 ppb row). Read
strips in the shade.
Table 2 Interfering substances
Ion or Substance
Concentration
Sulfide
>15 ppb
1
Selenium
> 1 ppm
Summary of Contents for Eclox
Page 1: ...2886888 Eclox Rapid Response Test Kit User Manual 01 2018 Edition 5 ...
Page 2: ...Visit us at www hach com ...
Page 6: ...Table of Contents Table of Contents Page 6 EcloxTOC fm ...
Page 14: ...14 ...
Page 24: ...24 ...
Page 46: ...46 ...
Page 52: ...52 ...
Page 110: ...110 ...
Page 113: ...113 ...
Page 114: ...Visit us at www hach com ...
Page 120: ...120 ...
Page 123: ...123 ...