68
Luminescent Bacteria Toxicity Test - Screening Luminescence Results
Sample: _______________________________________________________
Date: _______________
Time: _______________
Operator: ______________________________________________________
Comments:
Procedure:
1.
Add 1.0 mL of
reconstitution solution to the
reagent. Swirl to mix. Wait 5
minutes.
2.
Add 1.0 mL of stock
suspension to 10 mL of
Diluent. Shake to mix. Wait
15 minutes.
3.
Add one spoon of solid
NaCl to 7 mL of sample.
4.
Fill four test tubes with the
test suspension, 2% NaCl,
and sample in the order
shown in the table. Start the
timer after adding the first
sample volume. Set the
timer for 15 minutes.
5.
Push
ON
(green button) on
the luminometer. Go to the
Screening Luminescence
Menu.
6.
Measure test tube 1
(non-toxic reference).
7.
Measure the sample tubes
in the order 2, 3 and then 4.
Record the results in the
table.
Order
1st
2nd
3rd
%
inhib.
Rel.
units
Tube
Test
susp.
(mL)
2%
NaCl
(mL)
Sample
(mL)
Sample
Conc.
1
0.2
0.8
no
Non-toxic
reference
2
0.2
0.6
0.2
20%
3
0.2
0.3
0.5
50%
4
0.2
no
0.8
80%
% sample
% inhibition
Summary of Contents for Eclox
Page 1: ...2886888 Eclox Rapid Response Test Kit User Manual 01 2018 Edition 5 ...
Page 2: ...Visit us at www hach com ...
Page 6: ...Table of Contents Table of Contents Page 6 EcloxTOC fm ...
Page 14: ...14 ...
Page 24: ...24 ...
Page 46: ...46 ...
Page 52: ...52 ...
Page 110: ...110 ...
Page 113: ...113 ...
Page 114: ...Visit us at www hach com ...
Page 120: ...120 ...
Page 123: ...123 ...