Section 10 Luminescent Bacteria Toxicity
Test: Screening/LIMIT measure
The Luminescent Bacteria Toxicity (LBT) Test screening and LIMIT measure
procedures use the luminometer. Before doing either procedure:
•
Read
section 3.1, Overview on page 15
.
•
Do the procedures in
section 3.2, Prepare the luminometer for use on page 16
.
•
Print color copies of the Screening Luminescence Results Sheet to use in the field
(refer to page 68) from www.Hach.com.
This chapter describes the LBT screening luminescence procedure and LIMIT
measure procedure and contains the procedure steps.
The screening luminescence procedure and LIMIT measure procedure are used in
the field or in an emergency situation when a rapid assessment of the inhibitory
effects of a sample is necessary. The screening luminescence procedure and LIMIT
measure procedure are a simplified test procedure that uses the same reagents
according to ISO 11348 but at ambient conditions.
The LBT screening luminescence procedure and LIMIT measure procedure are
done the same with two exceptions:
•
Different options on the luminometer are selected to measure the sample
dilutions.
•
Different options on the luminometer are selected to show or send results.
•
A LIMIT value is set by the user using the luminometer for the LIMIT measure
procedure.
A column is added to the LIMIT measure test results that shows whether the
percent inhibition measured for each sample dilution is above or below the LIMIT
value (percent inhibition) set by the user on the luminometer.
Do the LBT screening luminescence procedure to do a toxicity screening measure.
Do the LBT LIMIT measure procedure to do a toxicity limit measure.
10.1 Overview
The screening luminescence procedure or LIMIT measure procedure is done to
identify if a sample is free of any inhibitory effects on the luminescent bacteria or, if
an inhibition is expected, to make an inhibitory or risk assessment of the sample.
Therefore, the user should measure dilutions of a sample and the percent inhibition
of the dilution steps to get more information about the severity of the inhibitory
effects.
In one run, the sample is measured in three different concentrations: 20% sample,
50% sample and 80% sample. The inhibitory effect of each sample dilution on the
Summary of Contents for Eclox
Page 1: ...2886888 Eclox Rapid Response Test Kit User Manual 01 2018 Edition 5 ...
Page 2: ...Visit us at www hach com ...
Page 6: ...Table of Contents Table of Contents Page 6 EcloxTOC fm ...
Page 14: ...14 ...
Page 24: ...24 ...
Page 46: ...46 ...
Page 52: ...52 ...
Page 110: ...110 ...
Page 113: ...113 ...
Page 114: ...Visit us at www hach com ...
Page 120: ...120 ...
Page 123: ...123 ...