9.2 pH calibration
1.
Push the
Power/Backlight key to set
the power to on. Remove
the sensor cap.
2.
Rinse the sensor and the
sensor cap with deionized
water. Dry with a no-lint
cloth.
3.
Push and hold the
Lock/Parameter key until
“pH” shows as the
measurement mode.
4.
Pour a pH 7.00 buffer to
the fill line of the sensor
cap.
5.
Put the sensor fully into
the sensor cap. The tester
reads the buffer value.
Air bubbles under the
probe tip when submerged
can cause slow
stabilization or error in
measurement. Shake the
tester from side to side to
remove air bubbles.
6.
Push the
Calibration/Settings key
to start the calibration.