108
11.13 Show or send previous results
Measured luminescent procedure values are stored by an indicator counter M1 to
Mx (refer to
Figure 12
). The counter starts with M1 every time the storage is erased
from the luminometer.
Figure 12 Example measurement luminescence results
To show previous results on the luminometer for the Luminescent Bacteria Test
(LBT), do the procedure in this section.
To send previous results to a computer:
Note:
At this stage, the results cannot be sent to the LUMISsoft 4.
1.
Do the steps in
Connect the luminometer to a computer on page 23
.
2.
Start LUMISsoft.
3.
In LUMISsoft, select Transfer, Options, Interface Protocol, Connect.
4.
Do the procedure in this section.
To send previous results to a printer, do the steps in
Connect the luminometer to a
printer on page 21
and then do the procedure in this section.
To show or send previous results saved on the luminometer:
1.
Push
ON
(green button) to apply power to the luminometer.
The luminometer does built-in tests that make sure that the electronics and
software are operating correctly.
2.
When the built-in tests are done, push
PROCEED
.
The Main Menu is shown.
3.
Select Luminescent Bacteria Test and push
ENTER
.
The LBT Main Menu is shown.
4.
Select Previous Results and push
ENTER
.
Summary of Contents for Eclox
Page 1: ...2886888 Eclox Rapid Response Test Kit User Manual 01 2018 Edition 5 ...
Page 2: ...Visit us at www hach com ...
Page 6: ...Table of Contents Table of Contents Page 6 EcloxTOC fm ...
Page 14: ...14 ...
Page 24: ...24 ...
Page 46: ...46 ...
Page 52: ...52 ...
Page 110: ...110 ...
Page 113: ...113 ...
Page 114: ...Visit us at www hach com ...
Page 120: ...120 ...
Page 123: ...123 ...