86
1.
Put 10 empty test tubes
into Row A of the
LUMIStherm.
2.
Set the 1.0 - 5.0 mL
pipette to 1.0 mL.
3.
Put the end of the pipette
into a clean pipette tip.
4.
Put the tip of the pipette
into the 2% NaCl solution
and slowly pull in 1.0 mL.
Slowly dispense the 2%
NaCl solution into the test
tube in position A9.
5.
Set the 1.0 - 5.0 mL
pipette to 1.5 mL.
6.
Put the tip of the pipette
into the 2% NaCl solution
and slowly pull in 1.5 mL.
7.
Slowly dispense the 2%
NaCl solution into the test
tube in position A8.
8.
Do steps
6
and
7
againto
put 1.5 mL of 2% NaCl
solution in each test tube in
positions A7, A6, A5, A4,
A3, A2 and A1.
Note:
Do not put 2% NaCl into
the test tube in position A10.